6 model 135/136 specifications @ 25 ˚c, Model 135/136 specifications @ 25 °c, Introduction – American Magnetics 135 & 136 Liquid Helium Level Instruments (CE-Marked) User Manual

Page 19

Advertising
background image

Rev. 3

7

Introduction

Specifications

1.6 Model 135/136 Specifications @ 25

°

C

Level Measurements

Resolution:

0.1%, 0.1 cm, or 0.1 in

Accuracy

a

:

±

0.5% of active sensor length

Linearity:

±

0.1%

Sensor Current:

75 mA nominal

Sensor Voltage:

approx. 70 VDC for 80" active sensor length

Operating Parameters

HI and LO Alarms:

0

% to 100% adjustable

HI/LO Alarm Relay Contact Ratings:

10 VA, 30 VAC or 60 VDC, 0.5 A

(normally open, closed on alarm)

Sample-and-Hold Period:

0.1 to 600.0 minutes or hours

A and B Control Setpoints:

0

% to 100% adjustable

Controller Output:

AC line voltage @ 1A max current

Fill Timer:

0.1 to 600.0 minutes or hours

0-10 Volt Analog Output

Integral Non-linearity:

±

0.012%

Resolution:

16 bits

Total Error:

±

1.1% for 0-10 V output

Voltage Drift (0-10 V):

100 ppm / °C

4-20 mA Analog Output @ 24 V

V

ext

Supply Range:

13-32 VDC (see

Appendix for diagram)

Integral Non-linearity:

±

0.012%

Resolution:

16 bits

Total Error:

±

0.25% for 4-20 mA output

Current Drift (4-20 mA):

75 ppm / °C

PSRR:

10

µ

A / V

Power Requirements

Primary

b

:

100-120 or 200-240 VAC

±

10%

50 - 60 Hz

Maximum Current:

50 VA for Model 135

1.3 A for Model 136

0RGHO 
0RGHO 
0RGHO 