4 commands for querying the configuration, Remote interface reference – American Magnetics 135 & 136 Liquid Helium Level Instruments (CE-Marked) User Manual

Page 46

Advertising
background image

34

Rev. 3

Remote Interface Reference

IEEE-488 Command Set Reference

the sampling interval to 0.1 minutes (or 0.1 hours if so configured, see
page 53). The SAVE command saves the HI, LO, A, B, INTERVAL,
LENGTH, and current remote units settings to permanent memory. These
settings are then recalled each time the power is turned off and then
reapplied to the instrument. If the configuration is changed from the front
panel, the settings are automatically saved to permanent memory.

4.2.4.4 Commands for querying the configuration

The LENGTH, HI, LO, A, B, and INTERVAL commands return the
current configuration of the instrument. Each return value is terminated
with <LF> and EOI.

Command: LENGTH

Function: Returns the sensor

length in the

current units

Returns: <value>

Command: HI

Function: Returns the HI

setpoint limit in the

current units

Returns: <value>

Command: LO

Function: Returns the LO

setpoint limit in the

current units

Returns: <value>

Command: A

Function: Returns the A

setpoint limit in the

current units

Returns: <value>

Command: B

Function: Returns the B

setpoint limit in the

current units

Returns: <value>

Command: INTERVAL

Function: Returns the

sampling interval in

the current units

Returns: <value>

0RGHO 

0RGHO 

Advertising