Installation – American Magnetics 135 & 136 Liquid Helium Level Instruments (CE-Marked) User Manual

Page 23

Advertising
background image

Rev. 3

11

Installation

Sensor Preparation

where R is the maximum allowable resistance (in ohms) for each lead wire
from the instrument to the sensor, and L is the active length of the
connected helium level sensor in inches. Please note that the maximum
allowable active sensor length for input line voltages less than 105 VAC or
210 VAC is 64 inches. Tables for active sensor length vs. lead wire distance
are provided below.

Minimum recommended gauge for stranded, tinned-copper lead wire

for input line voltages greater than or equal to 105 VAC or 210 VAC.

Distance

R=367

R=315

R=263

R=211

R=107

R=3.2

L=10”

L=20”

L=30”

L=40”

L=60”

L=80”

10 ft.

36 AWG

36 AWG

36 AWG

36 AWG

36 AWG

34 AWG

20 ft.

30 AWG

30 ft.

40 ft.

28 AWG

50 ft.

27 AWG

100 ft.

24 AWG

200 ft.

22 AWG

500 ft.

32 AWG 16 AWG

Minimum recommended gauge for stranded, tinned-copper lead wire

for input line voltages less than 105 VAC or 210 VAC.

Distance

R=284

R=232

R=180

R=128

R=24

L=10”

L=20”

L=30”

L=40”

L=60”

10 ft.

36 AWG

36 AWG

36 AWG

36 AWG

36 AWG

20 ft.

30 ft.

40 ft.

50 ft.

100 ft.

34 AWG

200 ft.

30 AWG

500 ft.

34 AWG

34 AWG

26 AWG