26

Rev. 3

Remote Interface Reference

Serial Command Set Reference

the sampling interval to 0.1 minutes (or 0.1 hours if so configured, see
page 53). The SAVE command saves the HI, LO, A, B, INTERVAL,
LENGTH, and current remote units settings to permanent memory. These
settings are then recalled each time the power is turned off and then
reapplied to the instrument. If the configuration is changed from the front
panel, the settings are automatically saved to permanent memory.

4.1.4.3 Commands for querying the configuration

The LENGTH, HI, LO, A, B, and INTERVAL commands return the
current configuration of the instrument. Each return value is terminated
with <CR><LF>.

Command: LENGTH

Function: Returns the active

sensor length in

the current units

Returns: <value>

<CR> <LF>

Command: HI

Function: Returns the HI

setpoint limit in the

current units

Returns: <value>

<CR> <LF>

Command: LO

Function: Returns the LO

setpoint limit in the

current units

Returns: <value>

<CR> <LF>

Command: A

Function: Returns the A

setpoint limit in the

current units

Returns: <value>

<CR> <LF>

Command: B

Function: Returns the B

setpoint limit in the

current units

Returns: <value>

<CR> <LF>

Command: INTERVAL

Function: Returns the

sampling interval in

the current units

Returns: <value>

<CR> <LF>