2 commands for configuring permanent memory, Remote interface reference – American Magnetics 135 & 136 Liquid Helium Level Instruments (CE-Marked) User Manual

Page 37

Advertising
background image

Rev. 3

25

Remote Interface Reference

Serial Command Set Reference

4.1.4.2 Commands for configuring permanent memory

The LENGTH command configures the active sensor length setting in the
current units. LENGTH=35.0 would configure the active sensor length to
35.0 units of centimeters or inches.

Note

The LENGTH=<value> command will only function if CM or INCH
are currently selected as the units of measurement. The LENGTH
command does not configure the Model 135/136 if the units of
measurement are PERCENT.

The HI and LO commands configure the high and low setpoint limit values
respectively. For example, HI=90.0 would configure the high setpoint limit
to 90.0 in whichever units of measurement last selected through the serial
interface. The A and B commands configure the upper limit and lower
limit of the control band, respectively. The HI, LO, A, and B commands are
compatible with the percent units selection.

The INTERVAL command sets the sampling interval used if the
instrument is set for sampling mode operation (as opposed to continuous
update) via the sampling mode toggle switch. INTERVAL=0.1 would set

Command: LENGTH=

<value>

Function: Configures the

active sensor

length

Returns: <CR> <LF>

Command: HI=

<value>

Function: Configures the HI

setpoint limit

Returns: <CR> <LF>

Command: LO=

<value>

Function: Configures the LO

setpoint limit

Returns: <CR> <LF>

Command: A=<value>

Function: Configures the A

setpoint (control

band upper limit)

Returns: <CR> <LF>

Command: B=<value>

Function: Configures the B

setpoint (control

band lower limit)

Returns: <CR> <LF>

Command: INTERVAL=

<value>

Function: Configures the

sampling interval

Returns: <CR> <LF>

Command: SAVE

Function: Saves the

configuration to

permanent

memory

Returns: <CR> <LF>