Remote Interface Reference

Serial Communication DIP Switch Settings

of only termination characters. Otherwise, the shared input/output
command buffer of the Model 135/136 may become corrupted. No other
form of flow control is supported.

4.1.3

Serial Communication DIP Switch Settings

The 8 DIP switches located on the rear panel of the Model 135/136 are
used to control various parameters of the serial interface. Switches 6
through 8 control the baud rate of the interface. Switches 3 through 5 are
unused. Switch 2 controls the echo feature and switch 1 controls the data
logger function. Each of these features is fully discussed below.

4.1.3.1 Baud rate control

The Model 135/136 baud rate is controlled by switches 6 through 8 of the
communication DIP switch on the rear panel. The unit is shipped with the
baud rate set at 9600. The switch settings for various baud rates are (on =
1 or the up position):

4.1.3.2 Echo function

The Model 135/136 has an echo feature which is enabled or disabled by
communication DIP switch 2. When the echo function is enabled, the
Model 135/136 will echo the incoming command characters back to the
transmitting device. The echo feature is useful when using an interactive
terminal program on a host computer for communicating with the Model
135/136 (however, most terminal programs support a local echo feature).
The settings are:

DIP switch

Baud rate

6

7

8

off

off

off

300

off

off

on

600

off

on

off

1200

off

on

on

2400

on

off

off

4800

on

off

on

9600

DIP switch 2

Function

on

Echo On

off

Echo Off

1

0

1 2 3 4 5 6 7 8

ON