A.3 auxiliary connector j2 pinout, Appendix – American Magnetics 135 & 136 Liquid Helium Level Instruments (CE-Marked) User Manual

Page 69

Advertising
background image

Rev. 3

57

Appendix

Auxiliary connector J2 pinout

A.3 Auxiliary Connector J2 Pinout

The HI level and LO level contacts are provided for external use by the
customer. When a HI or LO level condition exists, the respective contact
pairs are closed. All setpoints have 1/2 mm hysteresis, therefore the
respective contact pairs may “chatter” if the liquid sloshes, bubbles, etc.

The HI level and LO level contacts also provide positive indication of a
power-off condition. With a power-off condition, both the HI level and LO
level contacts will be closed, which is a state unique to the power-off
condition.

The following table provides the specifications for the contacts:

Pin

Function

1

4-20 mA current loop input

(optional feature)

2

4-20 mA current loop output

(optional feature)

3

0-10 VDC output

(optional feature)

4

0-10 VDC or 4-20 mA

common (optional feature)

5 & 6

Lo level relay contacts (dry)

7 & 8

Hi level relay contacts (dry)

9

Not used

Max switching VA

10

Max switching voltage

30 VAC or 60 VDC

Max switching current

0.5 A

Max continuous current

1.5 A