Error Codes

4.3 Error Codes

The Model 135/136 returns specific error codes for invalid commands and/
or arguments. If an error condition is returned, the command is not
processed and the configuration of the instrument is not modified. The
table below provides a list of error codes, their meaning, and any
associated limits. The error codes are common to both the serial and IEEE-
488 communication options.

Error Code

Meaning

Valid Range

1

LO setpoint out of range

0

LO

LENGTH

2

B setpoint out of range

0

B

<

A

3

A setpoint out of range

B < A

LENGTH

4

HI setpoint out of range

0

HI

LENGTH

5

Attempted to set or query for

LENGTH in PERCENT units mode

6

Invalid argument,

value out of

maximum calibration range

1 cm

value

203.2 cm

7

INTERVAL setting out of range

0

INTERVAL

600 min

8

Unrecognized command

9

Invalid argument, value was negative

or non-numeric

0RGHO 

0RGHO 