VEGA Conductive electrodes User Manual

Page 10

Advertising
background image

10

Conductive measuring system

Product description

Measuring data input, Ex-technical data

Signal circuits

in classification intrinsic safety EEx ia IIC

Max. values
- U

O

10,6 V

- I

K

8 mA

- P

20 mW

Characteristics

linear

Inner effective inductance

negligible

Inner effective capacitance

negligible

Max. permissible outer inductance

5 mH

Max. permissible outer capacitance

570 nF

The intrinsically safe circuits are reliably galvanically separated from the not-intrinsically
safe circuits up to a peak value of the nominal voltage of 375 V.
The intrinsically safe circuits of channel 1 and channel 2 are reliably galvanically isolated.

Relay outputs

Number

2

Mode

overfill protection (A) and dry run protection (B)
changeover for each output (channel) separately

Contact

1 spdt per output

Contact material

AgCdO and Au plated

Turn-on voltage

min. 10 mV
max. 250 V AC or 250 V DC

Switching current

min. 10 µA
max. 3 A AC or 1 A DC

Breaking capacity

max. 750 VA or 54 W

Transistor outputs

Number

2 (synchronously switching with relay outputs)

Galvanic isolation

floating

Turn-on voltage

U

B

= max. 36 V DC

Switching current

I

B

= max. 60 mA

Voltage loss on the transistor

U

CE

= approx. 1,5 V at I

B

= 60 mA

Blocking current

I

O

< 10 µA

Integration time

Fixed value

approx. 500 ms

Ambient conditions

Permissible ambient temperature

-20°C … +60°C

Storage and transport temperature

-40°C … +70°C

Advertising