2 mounting preparations – VEGA VEGASON 66 4 … 20 mA_HART four-wire User Manual

Page 13

Advertising
background image

1

3

2

100%

0%

4

Fig. 4: Version B, C, D - Measuring range (operating range) and max. measuring

distance

1

full (dead zone)

2

empty (max. measuring distance)

3

Measuring range

4

Reference plane

Gauge pressure in the vessel does not influence VEGASON

66. Low pressure or vacuum, however, damp the ultrasonic

pulses. This influences the measuring result, particularly if the

level is very low. With pressures under -0.2 bar (-20 kPa) you

should use a different measuring principle, e.g. radar or guided

microwave.

4.2 Mounting preparations

Version B is supplied in two parts: Transducer and electronics

housing.
Assemble the version as follows:
1 Loosen the screws on the fixing ring (4) with a hexagon

spanner (size 4), remove the transducer tube out of the

swivelling holder

2 Mount the flange
3 Insert the transducer tube in the requested length from

below into the swivelling holder

4 Fasten with screws (4); torque max. 10 Nm
5 Remove the plug from below out of the instrument housing

and plug it into the socket of the transducer tube

Pressure/Vacuum

Mounting - Version B

VEGASON 66 - 4 ... 20 mA/HART - four-wire

13

Mounting

28783

-EN

-060731