VEGA VEGAFLEX 66 (-20…+250°C) Profibus PA User Manual

Page 51

Advertising
background image

L

0

-3 mm (-0.118

")

3 mm (0.118

")

-50 mm (1.969

")

5 mm (0.197

")

0,15 m (5.906

"

)

0,03 m (1.181

"

)

0,02 m

(0.787

")

1

Fig. 24: Deviation VEGAFLEX 66 in coax version in water

1

Dead zone - no measurement possible in this area

L

Probe length

L

0

-3 mm (-0.118

")

3 mm (0.118

")

10 mm (0.394

")

5 mm (0.197

")

-5 mm (-0.197

")

0,15 m (5.906

"

)

0,1 m (3.937

"

)

0,15 m

(5.906

")

0,1 m

(3.937

")

1

1

Fig. 25: Deviation VEGAFLEX 66 in coax version in oil

1

Dead zone - no measurement possible in this area

L

Probe length

Influence of the ambient temperature to the sensor electronics

Temperature drift

0.03 %/10 K relating to the max. measuring range
or max. 0.3 %

Temperature drift - Digital output

3 mm/10 K relating to the max. measuring range or
max. 10 mm

Ambient conditions

Ambient, storage and transport temperature

-40 … +80 °C (-40 … +176 °F)