6 set up with vegadis 62, 1 connection, 2 adjust the sensor – VEGA VEGAWELL 52 4 … 20 mA_HART Pt 100 User Manual
Page 22: 6set up with vegadis 62
6
Set up with VEGADIS 62
6
.1 Connection
T
he VEGADIS 62 is an indicating and adjustment unit without external
energy for looping into 4 … 20 mA/HART circuits.
T
he parameter adjustment of the sensor is carried out via HART
communication. During the parameter adjustment, the VEGADIS 62
acts as a Secondary Master with respect to the sensor.
~
=
1
2
4
3
Fig. 13: Connection of VEGADIS 62 to the sensor
1
Sensor
2
VEGADIS 62
3
HART resistor > 150 Ω (with low impedance power supply necessary)
4
Voltage supply/Processing
T
he following adjustment volume of the connected HART sensor is
available:
l
M
in./Max. adjustment
l
zero/span adjustment (live adjustment)
l
D
amping
6
.2 Adjust the sensor
P
ush "OK" to reach the adjustment menu. Select with [↑] and [↓] the
submenu "Measurement" and confirm with "OK".
M
ove to the menu item "Unit". There the instrument unit of the sensor
is displayed, for example "bar".
M
ove further to the menu item "MB begin", there the measuring range
begin is displayed, for example the default setting 0 bar.
22
VEGAWELL
52 • 4 … 20
mA/HART Pt
6 S
et up with VEGADIS 62
35402
-EN
-111021