VEGA VEGAWELL 52 4 … 20 mA_HART Pt 100 User Manual
Page 33
Nominal range
Overload capacity, max.
pressure
Overload capacity, min.
pressure
0 … 70
psi
900
psi
0
psi
0 … 150
psi
1300
psi
0
psi
0 … 350
psi
1900
psi
0
psi
Output variable
O
utput signal
4 … 20
mA/HART
S
ignal resolution
1 µA
F
ault signal
<
3.6
mA; 20.5 mA; 22 mA; unchanged (adjustable
via PACTware)
M
ax. output current
22
mA
R
un-up time approx.
15
s
S
tep response time
≤
200
ms (ti: 0 s, 0 … 63 %)
HART
output value according to HART 5.0
-
PV (P
rimary Value)
P
ressure
Additional output parameter - temperature
integrated resistance thermometer
P
t 100 according to DIN EN 60751
R
ange
-50 … +100 °C (-58 … +212 °F)
R
esolution
1 °K
A
djustment external temperature transmitter
in VEGABOX 02
4 … 20
mA/HART according to -20 … +80 °C
(-4 … +176 °F)
Reference conditions and actuating variables (similar to DIN EN 60770-1)
R
eference conditions according to DIN EN 61298-1
-
T
emperature
+
15 … +25 °C (+59 … +77 °F)
-
R
elative humidity
45 … 75 %
-
A
ir pressure
860 … 1060
mbar/86 … 106 kPa (12.5 … 15.4 psi)
D
etermination of characteristics
L
imit point adjustment according to IEC 61298-2
C
haracterstic curve
L
inear
R
eference installation position
upright, diaphragm points downward
I
nfluence of the installation position
<
0.2
mbar/20 Pa (0.003 psig)
Deviation determined according to the limit point method according to IEC 60770
7
)
A
pplies to digital HART interface as well as to analogue current output 4 … 20 mA. Specifications
refer to the set span. Turn down (TD) is the relation nominal measuring range/set span.
D
eviation with version < 0.2 %
-
T
urn down 1 : 1 up to 5 : 1
<
0.2 %
7)
I
ncl. non-linearity, hysteresis and non-repeatability.
VEGAWELL
52 • 4 … 20
mA/HART Pt
33