T

he CERTEC

®

measuring cell is also equipped with a temperature

sensor Pt 100 in four-wire technology. The resistance value is
accessed via the wires of the suspension cable. The power supply or
the processing is carried out via a temperature transducer, e.g. in the
VEGABOX

02.

A

s a standard feature, the CERTEC

®

measuring cell is equipped with

a lateral, recessed seal.

I

nstruments with double seal have an additional seal in front.

T

wo-wire electronics 4 … 20 mA/HART for power supply and

measured value transmission over the same cable.

T

he supply voltage range can differ depending on the instrument

version.

T

he data for power supply are specified in chapter "Technical data".

3

.3 Operation

T

he instrument can be adjusted with the following adjustment media:

l

W

ith the external indicating and adjustment instrument VEGADIS

62

l

an adjustment software according to FDT/DTM standard, e.g.
PACT

ware and PC

l

W

ith a HART handheld

T

he type of adjustment and the adjustment options depend on the

selected adjustment component. The entered parameters are gen-
erally saved in the respective sensor, when adjusting with
PACT

ware™ and PC optionally also in the PC.

3

.4 Packaging, transport and storage

T

he device was protected by packaging during transport. Its capacity

to handle normal loads during transport is assured by a test according
to DIN EN 24180.

T

he packaging of standard instruments consists of environment-

friendly, recyclable cardboard. For special versions, PE foam or PE foil
is also used. Dispose of the packaging material via specialised
recycling companies.

T

ransport must be carried out under consideration of the notes on the

transport packaging. Nonobservance of these instructions can cause
damage to the device.

T

he delivery must be checked for completeness and possible transit

damage immediately at receipt. Ascertained transit damage or
concealed defects must be appropriately dealt with.

Seal concept