VEGA VEGAPULS 62 (≥ 2.0.0 - ≥ 4.0.0) 4 … 20 mA_HART four-wire Approval according to LPR radio standard User Manual
Page 69
Seal
Antenna impedance co-
ne
Process temperature (measured on the process
fitting)
FFKM (Kalrez 6230)
PTFE
-15 … +130 °C
(5 … +266 °F)
PEEK
-15 … +250 °C
(5 … +482 °F)
Graphite
Ceramic
-200 … +450 °C (-328 … +842 °F)
Vessel pressure - horn antenna
-
Antenna impedance cone PTFE
-1 … 40 bar (-100 … 4000 kPa/-14.5 … 580 psig)
-
Antenna impedance cone PP
-1 … 3 bar (-100 … 300 kPa/-14.5 … 43.5 psig)
-
Antenna impedance cone PEEK
-1 … 100 bar (-100 … 10000 kPa/-14.5 … 1450 psig)
-
Antenna impedance cone ceramic
-1 … 160 bar (-100 … 16000 kPa/-14.5 … 2320 psig)
Vessel pressure - parabolic antenna
-1 … 6 bar (-100 … 6000 kPa/-14.5 … 870 psig)
Vessel pressure relating to the flange nom-
inal stage
see supplementary instructions manual "Flanges
according to DIN-EN-ASME-JIS"
Vibration resistance
-
Horn antenna
4 g at 5 … 200 Hz according to EN60068-2-6
(vibration with resonance)
-
Parabolic antenna
1 g at 5 … 200 Hz according to EN60068-2-6
(vibration with resonance)
Shock resistance
-
Horn antenna
100 g, 6 ms according to EN60068-2-27 (mechan-
ical shock)
-
Parabolic antenna
25 g, 6 ms according to EN60068-2-27 (mechanical
shock)
Data on rinsing air connection
Pressure max.
6 bar (87.02 psig)
Air quantity without reflux valve, with
-
0.2 bar (2.9 psig)
2.0 m
3
/h
-
0.5 bar (7.25 psig)
3.5 m
3
/h
-
1 bar (14.5 psig)
4.5 m
3
/h
-
1.5 bar (21.76 psig)
5 m
3
/h
-
2 bar (29.0 psig)
5.5 m
3
/h
Thread
G⅛ A
Closure, with
-
Non-Ex
Dust protection cover of PE
-
Ex
Threaded plug of 316Ti
VEGAPULS
62 • Approval according to LPR radio standard
69
11 Supplement
41719
-EN
-120301