Fig. 8: VEGAMIP 62 - threaded version G1½ A

1

Polarisation marking

2

Instrument hexagon

3

Locknut

4

Process seal

1

Insert the supplied process seal (1) before screwing it in

2

To screw VEGAMIP 62 in, use the instrument hexagon (3) below
the housing.

Screw the instrument in and hand-screw it a flat wrench

3

Turn the sensor back (max. 180°) to reach the required orientation
of the polarisation marking (4)

4

Hold the instrument in this position and tighten with the counter nut
(2)

Caution:
The housing must not be used to screw the instrument in! Applying
tightening force can damage internal parts of the housing.

Note:
In the case of VEGAMIP 62 with NPT thread, the instrument seals in
the thread itself. Hence, no counter nut is necessary for these
versions.

Avoid long sockets in which the medium can remain and if possible,
mount VEGAMIP 62 front-flush. This applies mainly if buildup and dust
are expected.