VEGA VEGAMIP R62 Receiving unit - Relay User Manual

Page 32

Advertising
background image

Press the respective key until the indication is within the LED indication
strip. This means one or two LEDs on the LED indication strip will light
up.

C

B

A

2

3

1

1

1

Fig. 37: Indication strip

1

Indication (LED indication strip)

2

More sensitive - Key to the left

3

Less sensitive - Key to the right

A

Medium: Plastics (PP, PVC …), solvents

B

Medium: Building materials (sand, stones …)

C

Medium: Conductive substances (coal, ore, salt …)

Set the indication approximately to the middle of the LED indicating
board until the yellow relay control lamp changes status.

Depending on the vessel size, process and/or medium, you have to
adjust the sensitivity of VEGAMIP 62 to more or less sensitive. In
products with low signal damping, in small vessels or pipelines or with
buildup, it might be necessary to test the switching function.

By means of the LED indicating board, you can see the actual receive
level during adjustment.

When the indication moves to the right, the instrument will be more
insensitive, to the left more sensitive.

In applications with products providing good signal damping (eg
stones, ore, sand, cement) and where buildup and contamination is
expected, you should adjust the instrument to less sensitive. In this
case, the VEGAMIP 62 switches only with a relatively large signal
damping by the medium. Possible buildup does not influence the
measurement.

In case of buildup, press the key (to the right) until the flashing LEDs
are at the right edge of the LED indication strip. For strong buildup, you
can press the key several times to make the instrument less sensitive.
The display on the LED indication strip does not change, but the
instrument will become a little less sensitive with each pressing of the
key.

Receive level

Medium provides good
signal damping

32

VEGAMIP R

62 • - Relay

6 Setup

41912

-EN

-120228

Advertising