VEGA VEGAFLEX 81 Foundation Fieldbus Coax probe User Manual

Page 20

Advertising
background image

20

5 Connecting to power supply

VEGAFLEX 81 • Foundation Fieldbus

44215-EN-130910

1. Unscrew the housing cover
2. If a display and adjustment module is installed, remove it by turn-

ing it slightly to the left.

3. Loosen compression nut of the cable entry
4. Remove approx. 10 cm (4 in) of the cable mantle, strip approx.

1 cm (0.4 in) of insulation from the ends of the individual wires

5. Insert the cable into the sensor through the cable entry

Fig. 9: Connection steps 5 and 6 - Single chamber housing

Fig. 10: Connection steps 5 and 6 - Double chamber housing

6. Insert the wire ends into the terminals according to the wiring plan
Information:

Solid cores as well as flexible cores with wire end sleeves are insert-

ed directly into the terminal openings. In case of flexible cores without

end sleeves, press the terminal from above with a small screwdriver;

Advertising