6 Set up the sensor with the display and adjustment module

VEGAPULS SR 68 • Modbus and Levelmaster protocol

41371-EN-130624

Through this selection, the sensor is adapted perfectly to the product

and measurement reliability, particularly in products with poor reflec-

tive properties, is considerably increased.
Enter the requested parameters via the appropriate keys, save your

settings with [OK] and jump to the next menu item with the [ESC] and

the [->] key.

In addition to the medium, also the application, i.e. the measuring site,

can influence the measurement.
With this menu item, the sensor can be adapted to the applications.

The adjustment possibilities depend on the selection "Liquid" or "Bulk

solid" under "Medium".

The following options are available when "Liquid" is selected:

The selection "Standpipe" opens a new window in which the inner

diameter of the applied standpipe is entered.

The following features form the basis of the applications:
Storage tank:

Setup: large-volumed, upright cylindrical, spherical

Product speed: slow filling and emptying

Process/measurement conditions:

– Condensation

– Smooth product surface

– Max. requirement to the measurement accuracy

Storage tanke with product circulation:

Setup: large-volumed, upright cylindrical, spherical

Product speed: slow filling and emptying

Setup - Application