10 Supplement

VEGAPULS SR 68 • Modbus and Levelmaster protocol

41371-EN-130624

Installation reference conditions

Ʋ Min. distance to installations

> 200 mm (7.874 in)

Ʋ Reflector

Plane plate reflector

Ʋ False reflections

Biggest false signal, 20 dB smaller than the useful signal

Deviation with liquids

See following diagrams

1,0 m (3.28 ft)

10 mm (0.394 in)

2 mm (0.079 in)

- 2 mm (- 0.079 in)

- 10 mm (- 0.394 in)

0

1 2

3

Fig. 51: Deviation under reference conditions
1 Reference plane

2 Antenna edge

3 Recommended measuring range

Repeatability

≤ ±1 mm

Deviation with bulk solids

The values depend to a great extent on the application.

Binding specifications are thus not possible.

Deviation under EMC influence

≤ ±30 mm

Variables influencing measurement accuracy
Temperature drift - Digital output

±3 mm/10 K relating to the max. measuring range or

max. 10 mm

Additional deviation through electromag-

netic interference acc. to EN 61326

< ±50 mm

Characteristics and performance data
Measuring frequency

K-band (26 GHz technology)

Measuring cycle time approx.

700 ms

Step response time

1)

≤ 3 s

Beam angle

2)

Ʋ Horn antenna ø 40 mm (1.575 in)

20°

Ʋ Horn antenna ø 48 mm (1.89 in)

15°

Ʋ Horn antenna ø 75 mm (2.953 in)

10°

Ʋ Horn antenna ø 95 mm (3.74 in)

Ʋ Parabolic antenna

1)

Time span after a sudden measuring distance change by max. 0.5 m in liquid applications, max 2 m with bulk

solids applications, until the output signal has taken for the first time 90 % of the final value (IEC 61298-2).

2)

Outside the specified beam angle, the energy of the radar signal has a level which is reduced by 50 % (-3 dB)