11. CALIBRATION

Adjusts the scale for accurate weighing.
Calibrate the scale in the following cases.

‰

When the scale is first installed.

‰

When the scale has been moved.

‰

When the ambient environment has changed.

‰

For regular calibration.


11.1. Gravity acceleration correction

When the scale is first used or has been moved to another location, it should be
calibrated using a calibration weight.
But if a calibration weight is not available, the gravity acceleration correction will
compensate the scale. Change the gravity acceleration value stored in the scale to the
value of the area where the scale will be used. Refer to the gravity acceleration map at
the end of this manual.

Note: Gravity acceleration correction is not required when the scale will be

calibrated using a calibration weight at the place where it is to be used.


‰

Make sure that the scale is in the weighing

mode (displaying “kg” (“g”), “lb”, “oz” or “pcs”).


1. Press and hold the TARE switch until the

gravity acceleration value (in this example:
9.7985) appears, and release the switch.

2. Press the TARE switch to enter the

gravity acceleration value setting mode.

3. Change the displayed value using the following

switches.

UNITS To shift the blinking digit to the

right.

PRINT To increase the value of the

blinking digit by one.

4. Press the TARE switch. The setting value

is stored and “

end

” is displayed.

5. When calibration using a calibration weight is

to be performed, go to step 3 of “11.2.
Calibration using a weight”. To finish the
setting procedure, turn the power OFF.

Press and hold

(more than 4 seconds)

Release

Press

Press

Press