Btpr, Op-04 specifications – A&D Super Wash Down Scales SW-150KL User Manual

Page 37

Advertising
background image

35

‰

The installation procedure is the same as for OP-03. Refer to “13.2.1. Installing OP-03”.

‰

Set the function settings “

bp5

”, “

btpr

”, “

prt

”, “

5if

”, “

adr

” and “

aCk

” as

necessary.

‰

Before using OP-04, the function setting “

5if

” must be set to specify whether RS-

422 or RS-485 is used.
To connect more than one scale to a computer, set a different address to each
scale using the function setting “

adr

”.

13.3.2. OP-04 Specifications

RS-422/485 Specifications

Transmission system EIA RS-422 / 485
Transmission form

Asynchronous, bi-directional, half-duplex

Data format

Baud rate: 2400, 4800, 9600 bps

Data bits:

7 bits + parity 1bit (EVEN / ODD)

or 8 bits (non parity)

Start bit:

1 bit

Stop bit:

1 bit

Code:

ASCII

Terminator:

C

R

L

F

(C

R

:0Dh, L

F

:0Ah)

LSB

0

1

2

3

4

5

MSB

6

Circuit diagram






Stop bit

A-B

1 (-2V~-5V)

0 (2V~15V)

Start bit

Data bits

Parity bit

1 SDA (RS-422 Out)
2 SDB (RS-422 Out)
3 RDA (RS-422 In / RS-485 In-Out)
4 RDB (RS-422 In / RS-485 In-Out)
5 TRM (Termination)
6 SG (Signal ground)

7 SHLD (Shield)

100Ω

1 SDA (RS-422 Out)
2 SDB (RS-422 Out)
3 RDA (RS-422 In / RS-485 In-Out)
4 RDB (RS-422 In / RS-485 In-Out)
5 TRM (Termination)
6 SG (Signal ground)
7 SHLD (Shield)