Ack 1, Ack 0 – A&D Super Wash Down Scales SW-150KL User Manual

Page 39

Advertising
background image

37

Command examples (“_” indicates “Space” (20H).)

The examples below are for the function setting “

aCk 1

“ (Reply to commands).

The address ## = 23.
‰

Request a weight data.

Command

@ 2 3 Q C

R

L

F

Reply

@ 2 3 S T , + 0 0 1 2 . 3 4 5 _ k g C

R

L

F

Stable data

@ 2 3 U S , + 0 0 0 7 . 8 9 0 _ k g C

R

L

F

Unstable data

@ 2 3 O L , + 9 9 9 9 . 9 9 9 _ k g C

R

L

F

“E” display

‰

Zero the scale. (No reply for the function setting “

aCk 0

“.)

Command

@ 2 3 Z C

R

L

F

Reply

@ 2 3 Z C

R

L

F

The scale is in a condition that zero operation is possible.

‰

Send the current LO limit value.

Command

@ 2 3 ? L 1 C

R

L

F

Reply

@ 2 3 L 1 , + 0 0 0 2 0 0 C

R

L

F



















Advertising