Alarms – Burkert Type 8056 User Manual

Page 30

Advertising
background image

SE56

30

CODE

ANOMALIE DESCRIPTION

ACTION TO TAKE

0001

problem with watch-dog circuit

ADDRESS TO SERVICE

0002

Wrong work data configuration in eeprom

0004

Wrong safety data configuration in eeprom

0008 Defective

eeprom

0010

Defective keyboard (one or more keys are pushed
during the test)

0020

Power supply voltage (+3.3) is out of range

0040

Power supply voltage (+13) is too low (<10V)

0080

Power supply voltage (+13) is too high (>14V)

0200

Timeout calibration input (broken input circuit)

0400

Input stage gaining is out of range

Check the status cables connecting the transmitter to the sensor, the
grounding connections or the possible presence of strong and
anomalous noise sources

0800

Interruption on the coils circuit

Check the status cables connecting the transmitter to the sensor

0C00

Cumulative alarm 0800 + 0400

See single code

ALARMS

Messages, causes and corrective actions

Anomalies, codes and remedies

MESSAGE

CAUSE

ACTION TO TAKE

NO ALARMS

All works regularly

//

MAX ALARM

The flow rate is higher than the maximum threshold set

Check the maximum flow rate threshold set and
the process conditions

MIN ALARM

The flow rate is lower than the minimum threshold set

Check the minimum flow rate threshold set and
the process conditions

FLOW RATE >FS

The flow rate is higher than the full scale value set on the
instrument

Check the full scale value set on the instrument
and the process conditions

PULSE/FREQ>FS

The pulse generation output of the device is saturated and can
not generate the sufficient number of pulses

Set a bigger volume unit or, if the connected
counting device allows it, reduce the pulse
duration value

EMPTY PIPE

The measuring pipe is empty or the detection system has not
been properly calibrated

Check whether the pipe is empty, or repeat the
empty pipe calibration procedure

INPUT NOISY

The measure is strongly effected by external noise or the
connecting cable from transmitter to the sensor may be broken

Check the status cables connecting transmitter to
sensor, the devices grounding connections or the
possible presence of noise sources

EXCITATION

FAIL

The coils or the cable connecting the sensor are interrupted

Check the connecting cables to the sensor

CURR. LOOP

OPEN

The 4-20mA output on board or the optional one are not
correctly closed on a valid load

Verify the load is applied to the output (max 1000
Ÿ).

To disable the alarm, set the “mA VAL.FAULT”
value (menu alarm) to 0.

P.SUPPLY FAIL

Power supply different from that indicated on the label

Verify that the power supply is as indicated on the
label