User Unit Select

Engineering units not included in the standard selection
can be programmed into the manometer using the Units
Select register in the program mode. The value
programmed into this register is used to calculate the
desired unit of measure. An example of converting to
“Feet of H

2

O” will be shown in the following steps, using

the conversation factor of 1 PSI = 2.30894 FT H

2

O.

Keystroke

Display

1. Press the PRGM
key.

Top line reads “PROGRAM
MODE” and bottom line
reads “UNITS SELECT”.

2. Press the PRGM
key.

Top line reads “UNITS
SELECT” and bottom line
shows current engineering
units.

3. Press the up or
down arrow key until
“USER UNIT SELECT”
is displayed.

Top line reads “UNITS
SELECT”
Bottom line reads “USER
UNIT SELECT”.

4. Press the PRGM
key.

See note 1 at

bottom of this table.

Top line reads “VALUE=”.
Bottom line reads
“CHANGE?: YES”.

5. Press the PRGM
key to change the
value.

Top line reads “USER UNIT
VALUE”.

6. Start entering the
conversion factor by
pressing the up arrow
key until the first digit
reads 2.

Top line read

s “USER UNIT

VALUE”.
Bottom line reads
“20000000”.

7. Press the right arrow
key to enter the value
“2” and advance the
cursor to the next digit.

Cursor flashes to the right of
the “2”. Now numbers,
decimal point or blank space
can be entered.

8. Repeat step 6 and 7
until bottom line reads
2.30894

Bottom line reads “2.30894”.
Last digit “4” is blinking.

9. If an error is made
use the left arrow key
to move the cursor
back to the incorrect
digit. Then press up or

The digit that is corrected is
blinking.

Advertising