Re-calibration – 5 point edit – Meriam M2 – Series Smart Manometer User Manual

Page 26

Advertising
background image

20

Re-Calibration – 5 Point EDIT


To edit the calibration points for a 5 Point re-calibration
follow the steps below.
NOTE: If the factory default values are acceptable, skip this
section and proceed to the re-calibration 5-Point START
procedure.

Keystroke

Display

1. With unit OFF, press
and hold the MIN/MAX
key, turn the unit on
using the ON/OFF key,
then release

Top line reads “RE-CAL”.
Bottom line reads “EDIT”.

2. Press the PRGM key.

Top line reads “CAL POINT
1”. Bottom line displays the
cal point value.

3. Press the up/down
arrow keys to edit the
selected digit. Use the
left/right arrow keys to
change the cursor
position.

Note: For 0%

go directly to step 4.

Bottom line displays the cal
point value. The cursor
flashes at the first position
while the value is changed,
then moves to the right
position when the right
arrow key is pressed.

4. Press the right arrow
key while on the right
most digit to proceed.

Top line rea

ds “CAL POINT

2”. Bottom line displays the
cal point value.

5. Repeat steps 3 and 4
for CAL POINTS 2, 3, 4
and 5.

Top line reads “CAL POINT
2/3/4/5”. Bottom line
displays the cal point value.

6. After editing CAL
POINT 5 press the right
arrow key while on the
right most digit to
proceed.

Top line reads “RE-CAL”.
Bottom line reads “EDIT”.

7. To perform the 5-point
re-cal, press the up
arrow key until START is
displayed on the bottom
line.
OR
To exit without
performing the 5-point
re-cal press the left
arrow key

Top line reads “RE-CAL”.
Bottom line, “START”.
Continue with 5-Point Re-
calibration procedure at
step 3 on next page.
OR
Returns to Measure Mode.

Advertising