Data logging – Meriam M2 – Series Smart Manometer User Manual

Page 22

Advertising
background image

16

Data Logging

Data Logging can be used to record pressure
measurements. Two record modes are supported;
automatic and manual. In automatic mode, a pressure
value is captured every 5 seconds for 20 minutes,
resulting in 240 stored values. In manual mode, a
pressure value is captured each time the PRGM key is
pressed up to 240 values. The data collected during a
logging session can be viewed upon completion.

Keystroke

Display

1. From the Measure
Mode press the PRGM
key.

Top line reads “PROGRAM
MODE” and bottom line
reads “UNITS SELECT”.

2. Press the up arrow
key four times.

Bottom line reads “DATA
LOGGING”.

3. Press the PRGM
key.

Top line reads “DATA
LOGGING” and bottom line
reads “RECORD”.

4. Press the PRGM
key.

Top line reads “RECORD
MODE” and bottom line
reads “AUTO” or “MANUAL”.

5. Press the PRGM
key at AUTO to start
automatic logging or at
MANUAL to start
manual logging mode.

Top line reads
“RECORDING X” and
bottom line reads “XX.XX
UNITS”. AUTO records
value every 5 seconds.
Manual records value each
time PRGM key is pressed.

6. To stop recording
values at any time,
press the key.

Top line reads “DATA
LOGGING” and bottom line
reads “RECORD”.

7. To access recorded
values, press the 
key.

Top line reads “DATA
LOGGING” and bottom line
reads “VIEW”.

8. To view recorded
values, press the
PRGM key.

Top line reads “DATA LOG:

1” and bottom line displays

the value. Continue pressing
the  key to view all values.

9. Press the key 3
times.

Returns to Measure Mode.

The “Auto Shut-Off” timer is disabled for
Data Logging sessions. Be sure to end the

session to re-enable the Auto Shut-Off timer.

Advertising