Sampling

Introduction

Once you have selected a Particle Size and completed the Auto-Zero process
the Haz-Dust IV is ready to begin sampling.

Conditions

The following conditions must be met before starting the sampling process.

Condition...

For further Information

See Page...

The correct particle size must be selected.

3-2

The correct sampling inlet must be attached.

3-2

The correct date and time must be set.

2-5

The Auto-Zero process must be complete.

3-7

The alarm level must be set if sampling with
the alarm feature.

2-4

Sampling

Follow the steps in the table below to begin dust sampling.

Step Action

1

Select Sample Rate from the Special Functions Menu.

2

Select a sample interval using the table below.

Select...

For maximum sampling time of...

1 Second

6 Hours

2 Seconds

12 Hours

10 Seconds

60 Hours

Note: A sample is taken each second and averaged by the sample
interval time selected.

3

Choose the memory storage type using the table below.

To... Select...

Erase all previously recorded

data points in all locations.

Overwrite, then
Select Yes to confirm,

Note: Selecting No will
cancel sampling process
without effecting memory.

Add data points to the next

consecutive location.

Continuation.

Note: See page 3-13 for explanation of location codes.

Continued on next page