Battery Maintenance

Introduction

The battery pack is a 6.0 V NiMH rechargeable battery that can hold a charge
for up to 8 hours. It is important to check the battery periodically and
recharge when necessary.

Checking the
Battery

The battery status can be checked using the menu options on the Haz-Dust
IV. Use the following menu options to check the battery.

Step Action

1

Select Special Functions from the Main Menu.

2

Select Systems Options.

3

Select Extended Options.

4

Select Battery Status.

Result: The Battery Level Screen displays the charging level of
the unit’s battery in VDC.

5

If the charge level is...

Then...

6.2 VDC or higher

The battery is fully charged.

6.2 VDC or lower

The battery must be recharged.
See instructions on page 5-9.