1-4

Overview of the EPAM-7500

Ease of use

The user controls all functionality and programming using menus displayed
on a high contrast LCD.

A 24-hour rechargeable battery capacity.

Automatic clean air purging of sensor for increased stability and accuracy.

Internal temperature compensation for ambient use.

General
Information

The LCD displays real-time concentration in milligram per cubic meter

(mg/m

3

) in accordance with EPA or OSHA Reference Methods.

Statistical information of TWA, STEL, Max and Min levels can be viewed

instantly.

The Haz-Dust is calibrated using Arizona Road Dust (ARD) against NIOSH

method 0600 for Respirable dust with a + 10% accuracy.

The calibration of the Haz-Dust can be adjusted to compensate for changes

in particle composition and distribution.

Figure 1.4. Diagram showing breathing zones of Inhalable, Thoracic, and Respirable dust
particles.