Adjusting the flow rate – SKC Limited EPAM-7500 User Manual

Page 48

Advertising
background image

5-5

Adjusting the Flow Rate

Introduction

The flow rate must be adjusted when it does not equal 2.0 LPM.

Adjusting the
flow rate

Follow the steps in the table below to adjust the flow rate.

1

Press Settings from the main menu.

2

Press the button until you are on the last screen.

3

Press the Lock graphic to unlock the settings.

4

Press the white box to enter the security code.

5

Press HD7 on the keyboard.

6

Press the Enter button on the keyboard.

7

Press the Check Box next to the Force Pump On.

8

Press the or to adjust Pump to 2.0 LPM.

9

Record the Flow Rate.

10

Detach the airflow calibrator and calibration airflow tubing.

11

Press Green check button to accept changes.