5 um dust particulates – SKC Limited EPAM-7500 User Manual

Page 31

Advertising
background image

3-6

2.5 um Dust Particulates

Follow the steps in the table below to select PM-2.5 dust particulates.

Step

Action

1

Select Settings from the Main Menu.

2

Select Sample Type.

3

Select 2.5 um – S.

4

Select Green check mark button to accept Changes.

6

Insert the sampling inlet into the sensor head of the EPAM-7500.
Note: If also collecting concurrent 47mm EPA FRM Style filter
samples place a clean gravimetric filter in the filter cassette. The
flow rate should be checked each time a new gravimetric filter is
used. See page 5-5 for information on checking the flow rate.

7

Attach the filter cassette to the sensor head of the EPAM-7500.
See diagram below.

8

Turn to page 3-7 and follow the instructions to Manual-Zero the
EPAM-7500.