Real-Time Dust Monitoring Principles

Principles

The EPAM-7500 uses the principle of near-forward light scattering of an

infrared radiation to immediately and continuously measure the
concentration in mg/m

3

of airborne dust particles.

This principle utilizes an infrared light source positioned at a 90-degree

angle from a photo detector.

As the airborne particles enter the infrared beam, they scatter the light. The

amount of light received by the photo detector is directly proportional to the
aerosol concentration.

A unique signal processes internally and compensates for noise and drift.

This allows high resolution, low detection limits and excellent base line
stability.

Figure 1.5. Diagram showing the principle of near-forward light scattering used in the Haz-
Dust..

Figure 1.6. Graph illustrating the principle of near-forward light scattering.


0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

0

0.5

1

R

e

la

ti

v

e

O

p

ti

c

a

l

R

e

s

p

o

n

s

e

Gravimetric Filter Weight

Light Scattering vs. Gravimetric Filter

ARD

0.1

ARD = Arizona Road Dust