5-6

Battery Maintenance

Introduction

The battery pack is a 10.8-volt Li-Ion rechargeable battery. It is important to
check the battery periodically and recharge when necessary.

Checking the
Battery

The battery status can be checked from any screen on the EPAM-7500. The
graphical icon is on the bottom middle of the screen.

Recharging the
battery

Follow the steps in the table below to recharge the battery using the supplied
charger.

Note: If the battery is too low, the EPAM-7500 will turn off. Charge the
battery for 15 minutes before turning on.

Step

Action

1

Plug the battery charger into an electrical outlet.

2

Plug the battery charger into the battery charge jack on the top
panel of EPAM-7500, or into the supplied charging cord.

Results:

LED indicator on front of the EPAM-7500 should be Red.

When LED is green battery is fully charged.


Note: Recharging time is approximately 22 hours when the
instrument is not operating.

CAUTION:

Do not charge in a hazardous environment. Use only

the EDC approved charger designed for the EPAM-7500.