WaveLink Medium-Bandwidth (8-13 GHz) Differential Probe

Solder-in Tips: Prevent damage to the solder-in tips by carefully storing them in the plastic
protective storage case when not in use.

Positioner Tip: The positioner tips are very small and are fragile. When not in use, store in
the provided case.

Connecting the Platform/Cable Assembly, Amplifier Module, and
Interconnect Lead

Follow these instructions to properly connect the main probe components.

Connecting an Amplifier Module to a Platform/Cable Assembly

Attach an Amplifier Module to the Platform/Cable Assembly by aligning the connectors of the
module with the receptacles in the platform/cable assembly and pressing the two together. Be sure
to finger-tighten the assembly by rotating the threaded collar onto the module.

Connecting the amplifier module to the platform/cable assembly.

Do not use pliers or any other tools to tighten the collar.

Remove the Amplifier Module by loosening the threaded collar from the module and pulling the two
assemblies apart.

By design, the amplifier module works with the WL-PLINK platform/cable assembly and the Dxx30-
SI Lead, Dxx30-SP Lead, Dxx30-SMA-SMP Lead, or Dxx30-PT Tip.