11. Adjust Out 2 100%

a. Connect a digital voltmeter between J6-1

Out 2 and J6-2 Gnd.

b. Press the UP or DOWN arrow key (and

RAPID key if necessary) until the desired
output voltage is attained.

c. Press ENTER.

12. Setup Complete

a. Press ENTER.

Programmable Threshold Relay

The Threshold Relay can be programmed to
Close on Higher (if the total tension goes above)
or Close on Lower (if the total tension goes
below) a preset programmable threshold
between 0% and 102% of the total tension. The
dry relay contact is rated at 1 amp @ 24VDC,
250mA @ 220 VAC, or 500 mA @ 120 VAC.

A lamp may be connected between J7-3 NO (or
J7-1 NC) and J7-3 COM during setup, start up,
and/or normal operation to verify the relay is
operating properly.

To set the threshold value:

1. Press the SETUP & RAPID keys simultane-

ously. The display will show a message indi-
cating the current value of the threshold. The
Run Mode LED will continue to flash.

Threshold 000%

2. Press the UP or DOWN arrow key until the

desired threshold value is displayed.

3. Press ENTER to select the value.

4. Press the UP or DOWN arrow key to toggle

between the CLOSE ON LOWER or CLOSE
ON HIGHER prompt.

Close on Lower

21

Warner Electric • 800-825-9050

P-2012-2

5. Press ENTER to select the desired mode.

Note: The threshold may be set while the
web machinery is in operation. Although
the LCD will not be showing the tension,
Output #1, Output #2, and the serial output
will still be responding to tension changes.

If the threshold value or mode is reset with the
equipment operating, the new value will take
effect immediately.

Reset Tare to Zero

The tare value may be zeroed to compensate
for any offsets accumulated during normal oper-
ation. Press the SETUP and EXIT keys simulta-
neously to reset the tare to zero.

Note: Resetting tare to zero must be done
with no tension load applied.