The FM style load cell provides easy and conven-
ient mounting to the roll that is being measured
(used with pillow blocks). The load cell is a strain
gauge style unit that is ideal for heavy tension
applications. It can be mounted regardless of ori-
entation, but has to work in compression. Only
the perpendicular force (resultant) is measured by
the load cell. The perpendicular force can be at a
maximum permitted angle of +/- 30 degrees.
Correct load cell sizing must be adhered to so
potential overload forces do not damage cells.

Specifications:

Load

Ratings

N

100 250

500 1000 2500 5000 10000

(Lbs.)

(22) (56) (112) (225) (562) (1124) (2248)

Size

01

01

01

01

01

01

02

Input Power

±12 to ±15 VDC
@ 45 MA ±5%

Output Signal

5VDC output at rated load

Temperature Range

0-70 degrees C
(32 -158 degree F)

Temperature drift

0.1 % of rating per degree C

Non-linearity and Repeatability

< 0.5%

Power Consumption

1 Watt

Cable

16 ft. provided with load cell

Maximum Load Ratings (See Figure 6)

Overload
120% of rated load
Compression Overload
150% of rated load
Horizontal Load
50% of rated load

FM Series Part Numbers

F M

0 1

-

1 0 0 0

-

A C

Model Size Load in N Amplifier built in

Alignment

Zero setting:
Green LED ON
above 0 VDC

Gain setting:
Green LED ON
below 5 VDC

Green LED

D-Sub connector

The sensor has been factory calibrated:

0VDC (No load)
5VDC (Rated load)

Two potentiometers and LED’s are located near
the “D” connector (See Figure 1) for visual indi-
cation. When used in the nominal range, both
LED’s are “ON”. Although the load cell has been
factory calibrated. The unit may be rescaled if
necessary. It is recommended that a qualified
technician do this procedure.

Attach a digital volt meter between the signal
lead (yellow) and 0VDC (ground). With no load
on the cell adjust the zero potentiometer for a
0VDC output. The zero setting LED should be
“OFF”. Any voltage above 0VDC, the zero LED
will turn “ON”.

Apply the mazimum load to the cell. Adjust the
gain potentiometer for a 5VDC output reading.
The gain LED should be “OFF” at a 5VDC output
level. Remove the load and insure the gain LED
is “ON”.

The zero and gain adjustments may need to be
repeated a couple of times to insure proper set-
ting.

Figure 1

Model Numbers/Part Numbers

FM01-100

6910-840-100

FM01-250

6910-840-102

FM01-500

6910-840-104

FM01-1000

6910-840-106

FM01-2500

6910-840-108

FM01-5000

6910-840-110

FM01-10K

6910-840-112