NIBP System Messages
9650-001355-01 Rev. H
X Series Operator’s Guide
8-15
PRESS. LIMIT EXCEEDED
The cuff pressure has exceeded the allowable limit.
Restrict patient movement and check the hose for
kinks. Eliminate kinks or replace the hose, if
necessary.
TEMP OUT OF RANGE
The NIBP module is outside of its operating
temperature range. Move unit to proper operating
temperature and reattempt the NIBP measurement.
WEAK PULSE
The patient’s pulse is too weak to obtain an NIBP
measurement or the cuff requires adjustment.
System Message
Cause