Observing the 12-lead waveform traces, Lead interpretive analysis – ZOLL X Series Monitor Defibrillator Rev H User Manual

Page 182

Advertising
background image

C

HAPTER

14 12-L

EAD

ECG I

NTERPRETIVE

A

NALYSIS

14-8

www.zoll.com

9650-001355-01 Rev. H

Observing the 12-Lead Waveform Traces

To observe the 12-Lead waveform traces, press

. The screen displays all twelve waveform

traces, with the size displayed above the waveform traces:

12-Lead Interpretive Analysis

After observing the patient’s ECG and determining that all 12-lead traces display correctly, you
can initiate 12-Lead Interpretive Analysis (12-Lead Interpretive Analysis must be enabled
through the Supervisor Menu).

Note:

12-Lead Interpretive Analysis works for only Adult patients.

To begin 12-Lead Interpretive Analysis, press the Acquire

quick access key (

). The

X Series unit displays the Acquiring 12-Lead Status bar as it collects 10 seconds of 12-Lead
ECG data:

12

06/06/2011 12:34:56

Adult

00:17:43

12

Exit

V6

V5

V4

V3

V2

V1

aVF

aVL

aVR

III

II

1 mV/cm

I

12

HR bpm

80

NIBP mmHg

121

79

(96)

EtCO2 mmHg

12

BR

38

SpO2 %

97

SpMet 20.0

T1 ºF

98.6

12