Using see-thru cpr, Examples – ZOLL X Series Monitor Defibrillator Rev H User Manual

Page 234

Advertising
background image

C

HAPTER

20

S

EE

-T

HRU

CPR

20–2

www.zoll.com

9650-001355-01 Rev. H

of asystole or low amplitude PEA, the residual artifact seen after filtering may look like fine
ventricular fibrillation.

Because the filtered ECG signal may contain residual chest compression and/or filtering
artifacts, a rescuer should always follow the standard procedure of stopping CPR to assess the
patient’s ECG rhythm before determining treatment
.

Using See-Thru CPR

To use See-Thru CPR

The X Series unit must be monitoring CPR.

ZOLL OneStep CPR pads, OneStep Complete pads, or CPR-D-padz must be attached to the
unit.

When chest compressions begin, the X Series unit automatically starts filtering the CPR artifact
after detecting the first 3 to 6 compressions. The filtered ECG, with the label “FIL,” may be
displayed on the second or third trace (by selecting FILT ECG in the Trace2 or Trace3 menu).

See-Thru CPR filtering continues as long as the OneStep CPR, OneStep Complete or
CPR-D-padz detect compressions and patient impedance is valid. When no compressions are
detected or one of the conditions noted above occurs, See-Thru CPR filtering stops, and
unfiltered ECG signals are displayed. When compressions resume, filtering automatically
restarts after 3 to 6 chest compressions.

Note:

There is a delay of approximately 1/16 second between the See-Thru CPR waveform
and the Trace 1 ECG waveform.

Examples

The following examples show the effects of See-Thru CPR filtering on ECG signals
contaminated with CPR artifacts.

Each example includes:

ECG signal with CPR artifact.

ECG signal after the See-Thru CPR filter has removed CPR artifact.

Indication of the period during which See-Thru CPR is active.

CPR signal to show when CPR activity occurred.

Advertising