2 charge defibrillator – ZOLL X Series Monitor Defibrillator Rev H User Manual

Page 198

Advertising
background image

C

HAPTER

15

M

ANUAL

D

EFIBRILLATION

15–8

www.zoll.com

9650-001355-01 Rev. H

The selected energy level is shown on the display.

2 Charge Defibrillator

Press the CHARGE button on the front panel.

To increase or decrease the selected energy after you have pressed the CHARGE button, use
the defibrillator Energy Select arrows on the front panel.

Caution

Changing the selected energy while the unit is charging or charged causes the defibrillator to
disarm itself. Press the CHARGE button again to charge the unit to the newly selected energy
level.

A charging message displays at the bottom of the screen, and a distinctive charging tone sounds
indicating that the unit is charging.

The energy range bar graph on the right side of the display highlights the charge level until it
reaches the selected energy. When the unit is fully charged, the tone changes to a continuous
charge ready tone, the highlighted energy bar graph includes the selected energy, and the shock
button lights up.

SHOCK

1

2

3

200

J

120

J

Selected Energy

1

2

4

5

6

8

9

10

15

20

30

50

120

70

150

200

3

7

85

100