Cleaning and Reuse of Sensors

Reusable sensors can be cleaned as follows:

Disconnect the sensor from the patient cable, if appropriate. Wipe the entire sensor clean with a
70% isopropyl alcohol moistened pad. Allow the sensor to air dry before returning it to use.

Connecting the SpO

2

Sensor

To connect the SpO

2

sensor to the X Series unit:

1. When using a sensor extension cable, inspect the cable before use. Replace the cable if it

shows any signs of wear, breakage, or fraying. Plug the sensor extension cable into the SpO

2

receptacle on the side of the X Series unit:

Figure 10-3 Connecting the SpO

2

Sensor to the X Series Unit

Displaying SpO

2

, SpCO, and SpMet Measurements

When the connection is made between the SpO

2

sensor and the unit, the message,

INITIALIZING, appears in the SpO

2

display. After a brief delay, the unit displays the SpO

2

measurement. If SpCO and SpMet are installed, the SpCO and SpMet measurements will
alternate every two seconds underneath the SpO

2

display.

If the message, SENSOR FAILURE, appears, the sensor is either incompatible with the X Series
unit, or it is not working, and you will need to replace the sensor.

A blip bar appears on the right side of the SpO

2

numeric display window. The blip bar indicates

the pulse derived from the plethysmograph normalized waveform.