Energy Select
The default energy selections for adult patients are:
Shock 1 - 120 joules
Shock 2 - 150 joules
Shock 3 - 200 joules
The default energy selections for pediatric patients are:
Shock 1 - 50 joules
Shock 2 - 70 joules
Shock 3 - 85 joules
Note:
Pediatric defibrillator energy levels should be selected based on site specific protocols.
Note:
If the X Series unit has been configured to begin CPR upon start up, it will
automatically begin with the CPR interval.