„ FP3500-T11/FP3600-T11/FP3650-T41 (Type AC)
‹Power Cable Connecting

‹Connecting the FP Power Cord

(1) Be sure that the FP’s power cord is not plugged in to the power supply.
(2) Remove the Terminal Strip’s clear plastic cover.
(3) Remove the screws from the two (2) terminals (L,N) and FG (Frame Ground)

Terminal, position the Ring Terminals and reattach the screws. (Check each
wire to make sure the connections are correct)

(4) Reattach the Terminal Strip’s clear plastic cover.

AC cable

Grounding Wire

Power Cord

Double insulation wire
1.25~2.0 mm

2

(16-14AWG)

1.25~2.0 mm

2

(16-14AWG)

Recommended
Ring terminal

*1

*1.

In order to prevent a short circuit caused by loose screws, make sure to
use a crimp-type terminal with insulating sleeve.

V2-MS3 compatible
(J.S.T. Mfg. Co.,Ltd).

Over ø3.2 mm[0.13 in.]

V2-P4 compatible
(J.S.T. Mfg. Co.,Ltd).

Over ø4.3 mm[0.17 in.]

• The torque required to tighten these screws are as follows:

Terminal Block: 0.5 to 0.6N•m.

FG (Frame Ground) Terminal: 0.6 to 0.7 N•m.

Under 6.0 mm
[0.24 in.]

Under 7.0 mm
[0.28 in.]

DON’T

CONNECT

FG (Frame Ground) Terminal