<12. TROUBLESHOOTING>

12-3

IM 01F06A00-01EN

12.4 Output is Indicated at Zero Flow

Is fluid flowing?

Are parameters configured
correctly at operating
conditions?

Stop flow.

No

Yes

Configure the parameters correctly.

No

Are the load resistance
and supply voltage within
the tolerance limits?

Yes

Adjust to within the tolerance limits.

Yes

No

Yes

Is digitalYEWFLO
properly grounded?

No

Ground digitalYEWFLO.

No

Yes

Does low cut adjust?

Adjust to low cut, refer to Section
10.2 "Adjustment for Manual Mode"

Does the tuning execute?

No

Execute the tuning, refer to Section
10.2 "Adjustment for Manual Mode"

Yes

Does high vabrations
occurs in pipeline?

No

Eliminate vibration noise using pipe
support.

Yes

Are pulsation produced?
Check the pipeline conditions,
refer to Chapter 3 "INSTALLATION."

F1203.ai

Yes

Was a faulty area
found with selfdiagnosis?

No

Check for recovery measures, refer to Section
6.5 "Self-Diagnostic (Error Code List)."

‡,IDEXLOWLQLQGLFDWRULVDWWDFKHG

check the display of the error code.

‡&RQQHFWDKDQGKHOGWHUPLQDODQG

check self-diagnostic.