<10. OPERATION>

10-2

IM 01F06A00-01EN

10.1.4 Totalizer Start and Totalizer Reset

When using the Totalizer Function, the start setup
should be done.
(1) Start operation using BT200
Enter to B40(TOTAL START), and move the

video bar to “EXECUTE”. Push “ENTER” key
at 2 times.

(2) Start operation using indicator
Enter to “Setting mode”, move to B40 of

parameter number, and enter to “01” of data
number.

Refer to Section 5.4 “Setting Mode.”
Totalized value can be reset using the indicator or
BT200.
(1) Reset operation using BT200
Enter to B42 (TOTAL RESET), and move the

video bar to “EXECUTE”. Push “ENTER” key at
2 times.

(2) Reset operation using indicator
Enter to “Setting mode”, move to B42 of

parameter number, and enter to “01” of data
Number.

Refer to Section 5.4 “Setting Mode.”

10.1.5 Setting of Pulse Output (Scaling)

Pulse output are constructed by two units, that are
“Scaled pulse and Unscaled Pulse”.

(1) Scaled Pulse

When SCALED PULSE is selected in B20, set
fl owrate per one pulse output. Rate unit is linking to
the fl ow unit.

(2) Unscaled Pulse

When UNSCALED PULSE is selected in B20, it
outputs the pulse calculated by following formula.
The formula for output pulse number is as follows.
Output pulse number per one second = vortex
number per one second / PULSE RATE set
number.
Refer to Section 11.6 “Flow Calculation.”

● Pulse Rate setting

Pulse rate setting is settable by “B21:PULSE
RATE”.

10.1.6 Setting of Burnout Switch

digitalYEWFLO is equipped with a CPU error
burnout function used to set the output direction
upon CPU error, and a sensor burnout function
that sets the direction of the output in the event of
burnout of the temperature sensor. When factory-
shipment under normal conditions, the output of
both CPU error burnout and sensor burnout are
set to HIGH, but if option code /C1 is specifi ed,
the CPU error burnout is set to LOW(-2.5% below)
output, and sensor burnout is set to LOW(-2.5%
below) output, respectively. The setting of the
direction of output from burnout can be changed.
To change the direction of output arising from
burnout, switch the setting pin on the CPU
assembly (Refer to Table 10.1).

Table 10.1

Output Setting Pin for Burnout

Pin

position

CPU error

burnout

direction

CPU error

burnout

output

Remark

H

L

HIGH

110% or more

(21.6mA DC)

Set to HIGH

before shipment.

H

L

LOW

-2.5% or less

(3.6mA DC)

Set to LOW for

option code /C1.

TP2

COM

HHT

P

F1002.ai

Pin position of Burnout Switch

Pin position of
Write Protect Switch

Figure 10.2 Pin position of Burnout and Write

Protect Switch