<12. TROUBLESHOOTING>

12-4

IM 01F06A00-01EN

12.5 Multi-Variable Type (/MV)

Are the power polarities
Correct?

Are the load resistance
and supply voltage within
the tolerance limits?

Check the polarities.

No

Yes

No

Are the sensor
connected correctly?

Yes

Adjust to within the tolerance limits.

Yes

Yes

No

Are the parameters
confiured correctly at
operational conditions?

Configure the parameters correctly.

No

Refer to other trouble shooting.

F1204.ai

Yes

Was a faulty area
found with self-diagnostic?

No

Check for recovery measures, refer to
Section 6.5 "Self-Diagnostic (Error Code List)."

‡,IDEXLOWLQLQGLFDWRULVDWWDFKHG

check the display of the error
code.

‡&RQQHFWDKDQGKHOGWHUPLQDODQG

check self-diagnostic.

Check the sensor.

Advertising