4 other operations for the hart configuration tool, Important – Yokogawa digitalYEWFLO (DY) User Manual

Page 65

Advertising
background image

<8. OPERATION VIA HART CONFIGURATION TOOL (HART 5)>

8-5

IM 01F06A00-01EN

NOTE

When using a Fieldmate, close the main display
of FieldMate.

b. Restart the device
Turn off the power to the device, and turn it on.

IMPORTANT

New protocol revision is applied only after having
performed restart of the device.

NOTE

A new HART revision number is displayed on the
indicator after restart the device.

(5) Confi rmation of the protocol revision

number

Confi rming the new protocol revision
a. Restart the HART confi guration tool

NOTE

When execute the other parameter confi guration
or setting change, execute after restart the
confi guration tool.

b.

Confi rm the new HART protocol revision
number

Callup the Universal rev parameter,
and confi rm that the new HART revision
number is displayed.

• Procedure to call up the Universal rev.

parameter.

[Root Menu]

 Review  Review1 

Universal rev

5: HART protocol revision 5
7: HART protocol revision 7

8.10.4 Other Operations for the HART

Confi guration Tool

Regarding other operations for the HART
Confi guration Tool, refer to the HART Confi guration
Tool operations manual.