5 end processing method of dyc signal cable, 1 for vortex flowmeter (dy-n), End processing method of dyc signal cable -6 4.5.1 – Yokogawa digitalYEWFLO (DY) User Manual

Page 25: For vortex flowmeter (dy-n) -6, Caution, Wiring

Advertising
background image

<4. WIRING>

4-6

IM 01F06A00-01EN

4.5 End Processing Method of DYC Signal Cable

4.5.1 For Vortex Flowmeter (DY-N)

1

2

3

4

5

6

7

8

9

Do not short-circuit the conductive layer and the
terminals (A, B, C and T*

1

).

Slide FEP (fluorinated ethylene propylene) tubing
over the twisted inner and outer drain wires C until
the tubing cannot be slid any further, and then cut
off the tubing leaving 5 mm (0.2 in.) of the stranded
drain wires exposed.

Slide heat shrinkable tubing over the cable end so
that the tubing covers the braided shield and
overlaps both the polyethylene jacket and loose
wires A, B, C, and T*

1

.

Slide a short piece of heat shrinkable tubing over
each of wires A, B, C, and T*

1

. Install a crimp-on

terminal lug at the tip of each wire. Crimp and
solder each lug.

Slide each short piece of heat shrinkable tubing
over the crimp sleeve. Heat all pieces of heat
shrinkable tubing with a heat blower or dryer.

Attach an identification label to the end of the cable.

Description

Figure

Unit : mm

(approx. inches)

F0407.ai

Strip off the outer polyethylene jacket, outer braided
shield and inner jacket, and inner braided shield as
per the dimensions below.

Strip off the black conductive layer convering two
wires completely, as per the dimensions below.
Twist each of the conductor and drain wires so that
there are no free strands.

Strip off about 5 mm (0.2 in.) of insulation for each
of wires A, B, and T*

1

, and twist the strands of each

wire. Twist the inner and outer drain wires together.

(*1): Only for /MV

T*

1

(Yellow)

90 (3.5)

5 (0.2) 10 (0.4)

5 (0.2)

Conductive Layer (Black)

Conductive Layer (Black)

A (Red)

T*

1

(Yellow)

40 (1.6)5 (0.2)

50 (2.0)
60 (2.4)

B (White)

3 (0.1)

or less

Drain wires

T*

1

(Yellow)

5 (0.2)

5 (0.2)

5 (0.2)

B (White)

A (Red)

C (Black)

C(Black)

Heat Shrinkable Tubing

Crimp and Solder Here

Lug tip

Heat Shrinkable Tubing

Heat Shrinkable Tubing

T

*

1

(Yellow)

A (Red)

B (White)

5 (0.2)

FEP Insulation Tubing

(Black)

C (Black)

T

*

1

(Yellow)

A (Red)

B (White)

10

NOTE

Check that the insulation resistance between each
wire including the inner shield is 10M or greater at
500V DC. Ensure that both ends of the wires are
disconnected (open-circuited) during the check.

NOTE

In case that the cable end fi nish
parts assembly is necessary
after delivery, contact your
nearest Yokogawa sales offi ce
or the sales representative
from which you purchased the
product.

CAUTION

Do not touch the '' conductive layer"
(black area covering the signal cables A
and B) to the converter case, terminal,
and other leadwires. If it is touched,
operation of the converter may be
incorrect. When the cable is terminated,
remove the conductive layer properly.

(B)

(C)

White

Black

(A)

Red

Unit : mm

(approx. inches)

F0408.ai

60 (2.36)

80 (3.15)

20 (0.79)

70 (2.76)

Yellow

*1

(T) 50 (1.97)

Figure 4.7

End Processing Method of DYC Signal Cable for Detector