11-6-2. On-line impedance checks

This check can be applied to a sensor with liquid earth electrode. In using a sensor having no liquid

earth electrode, set impedance check off on service code 3 and 4.

T5.3.1E.eps

Code Display

Function

Function detail

X

Y

Z Ddefault values

Parameter specific functions

03

*Z1.CHK

Impedance check 1

Low

0

1.1.1

High

1

High

Temp comp off

0

Temp comp on

1

On

Imp check off

0

Imp check on

1

On

*Z.L.x

V

Low impedance limit

Press NO to step through choice of units.

1 M

7

x=None, K, M or G

press YES to select units, then use the

, ENT keys to set the value

*Z.H.x

V

High impedance limit

Press NO to step through choice of units.

1 G

7

x=None, K, M or G

press YES to select units, then use the

, ENT keys to set the value

04 *Z2.CHK

Impedance

check

2

Low

0

0.0.1 Low

High

1

Temp comp off

0

Off

Temp comp on

1

Imp check off

0

imp check on

1

On

*Z.L.x

V

Low impedance limit

Press NO to step through choice of units.

100

7

x=None, K, M or G

press YES to select units, then use the

, ENT keys to set the value

*Z.H.x

V

High impedance limit

Press NO to step through choice of units.

200

7

x=None, K, M or G

press YES to select units, then use the

, ENT keys to set the value

Yokogawa sensors with the liquid earth electrode include the PH8EFP, PH8ERP, PH8EHP,

OR8ERG, OR8EFG and FU20.

The liquid earth electrode is inside the sensor, and at the potential of the measured solution. The

liquid earth electrode is also used for an impedance check (to determine if a functional sensor

is connected). Be sure to connect the wire from the liquid earth electrode to terminal 14 on the

transmitter.

The impedance check detects when the glass electrode is broken, and also provides a check of

junction contamination and junction clogging. Such checks are qualitative rather than absolute , and

relatively loose high/low limits for "acceptable" values are usually set.

In the measurement in low conductivity water such as pure water, the impedance check does not

work. Set the impedance check function for the reference electrode to off in Service Code 04.

If the sensor doesn't have a liquid earth electrode, it's not possible to perform an automatic

impedance check to determine if the sensor is connected. Turn the functions of Service Codes 3 and

4 off when using such a sensor.

Yokogawa sensors without liquid earth terminals include the HA405, HA406, DPA405, DPAS405,

DPA406, HF405, HA485, DPA485 and DPAS485.